BRONKHORST®
Page14RS232interface 9.17.027
Value Meaning
1,2,8 generalerror
3.4 COMMUNICATI ONMESSAGES
Communicationmessages between FLOW‐BUSinterfacesandotherdevicesconsist of commandstringswithspecific
information.Thiscommand stringiseitherASCII(RS232)orBINARY. Basically thestring contains several information
bytes.ThroughRS232thesehexadecimalbytesareconvertedinASCII(e.g.:bytevalue0x0Ais"0A"inASCIIand
capital
lettersshould be used).MessagesviaRS232arepreceded bythe“:”character andterminatedwith"\r\n"(Carriage
return‐Line‐feed).
ThereareseveralCOMMANDSavailableintheFLOW‐BUSmessages.OnlycommandRD(04)andWR(01)arerequired
forallthestandardparameterreadingandwriting.A
RDcommandwillbeansweredwithaWRcommand,containing
thevalueaskedfororastatusmessage,containinganerrornumber.A WRcommandwillbeansweredwithastatus
message,containinganerrornumber(iferrornumber=0,thanWRcommandwasOK).
SCIIcharacter:hashexadecimalvalue:3A
ASCIIcharacter‘\r’hashexadecimalvalue:0D
ASCIIcharacter‘\n’hashexadecimalvalue:0A
3.4.1 Communicationcommands
Communicationcommands
Command Description
00 Statusmessage
01 Sendparameterwithdestinationaddress,willbeansweredwithtype00command
02 Sendparameterwithdestinationaddress,nostatusrequested
03 Sendparameterwithsourceaddress,nostatusrequested
04 Requestparameter,willbeansweredwithtype02or00command
06 Stopprocess
07 Startprocess
08 Claimprocess
09 Unclaimprocess
Toaccessaspecificparameteryouneedtoknowthefollowingpoints.
Nodeaddress eachFLOW‐BUSdeviceisconnectedtoaspecificnodeaddressinthesystem.
Processnumber eachdevice(node)consistsofseveralprocesses.
Parameternumber(FBnr) eachprocessconsistsofseveralparameters.
Parametertype eachparametercanbeofadifferenttypeandvalue.
Forparametersnumbersandvaluesseetables“parameterproperties”and“parametervalues”inthis
manual.