BRONKHORST®
Page24RS232interface 9.17.027
N 10 Parameterindex16(return) Parameternumber31
Capacityunit
21 01 00000001
P 01 Process1
22 7F 01111111
T 60 Parametertype‘string’
P 1F Parameternumber(FBnr.)31– Capacityunit
23 07 00001110
07 Stringlength7
24 71 01110001
C 00 Parameternotchained(return)
Parameterindex17(return)
Process1
Parameternumber17
Fluidname
T 60 Parametertype‘string’(return)
N 11 Parameterindex17(return)
25 01 00000001 P 01 Process1
26 71 01110001
T 60 Parametertype‘string’
P 11 Parameternumber(FBnr.)17– Fluidname
27 0AStringlength10
29 \n LineFeed
28 \rCarriageReturn End
TotalString :1A0304F1EC7163006D71660001AE0120CF014DF0017F077101710A\r\n
Answerbynode3
Nr Byte Layout Description Descriptionperblock
0 :
Start
1 37
Numberofbyteswhichdo follow:65bytes
2 03
Node3
3 02
Commandwrite
4 F1 11110001
C 80 Processchained
Process113
P 71 Process113(receivingprocess)
5 EC 11101100
C 80 Parameterchained
Parameterindex12
Serialnumber
T 60 Parametertype‘string’
N 0C Parameterindex12
6 00
Lengthoftheanswer10Bytes
7‐26
4D363231323334354100
ParametervalueconvertedfromhextoASCII:M6212345A
27 6D 01101101
C 00 Processnotchained
Parameterindex13
USERTAG
T 60 Parametertype‘string’
N 0D Parameterindex13
28 00
Stringlength00,lengthnotdefined
29‐36
5553455254414700
ParametervalueconvertedfromhextoASCII,thevaluesdo
read:USERTAG
37 01 00000001
C 00 Processnotchained
Process1
P 01 Process1(receivingprocess)
38 AE 10101110
C 80 Parameterchained
Parameterindex14
Measure
T 20 Parametertype’integer’
N 0E Parameterindex14
39 1C
Parametervalueis:1CD8(hex)
MeasureValueis:7384(dec)
40 D8
41 CF 11001111
C 80 Parameterchained
Parameterindex15
Capacity
T 40 Parametertype‘float’
N 0F Parameterindex15
42 3F
3F800000
ParameterValueinIEEE‐floatingpointnotation,32‐bitsingle
precision
Parametervalueconvertedfromfloattodecimal,thevalues
reads:1.0
43 80
44 00
45 00
Commentaires sur ces manuels