Bronkhorst DeviceNet slave interface Manuel d'utilisateur Page 15

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 43
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 14
BRONKHORST®
Page15DeviceNetinterface 9.17.026
0x31 0x73 CounterConvergence
factor
0x0E,0x10 REAL 104,10 0.0….1.0
(miniCORIFLOWonly)
0x31 0x74 CounterController
Gain
0x0E,0x10 REAL 104,11 (miniCORIFLOWonly)
0x31 0x75 Reset 0x10 USINT 115,8 0=NoAction
1=Resetcounter
3=Resetcounter
4=Resetandstopcounter
1
DependsonthevalueassignedtotheDataTypeattribute.Ifthevalueofthisattributeis0xC3,the
selecteddatatypeisInteger.Ifthevalueofthisattributeis0xCA,theselecteddatatypewillbethe
IEEE754singleprecisionfloatingpoint.
ANALOGSENSOROBJECTSERVICES
ANALOGSENSOROBJECT SERVICECODE SERVICENAME SERVICEDESCRIPTION
0x31 0x0E Get_Attribute_Single Returnsthecontentsofthespecifiedattribute
0x31 0x10 Set_Attribute_Single Modifiesanattributevalue
0x31 0x4B ZeroAdjust StartAutozero
3.3.6 SAnalogActuatorObject
ClassCode32HEX
TheSAnalogActuatorObjectmodelstheinterfacetoaphysicalactuatorinadevice.
ANALOGACTUATOROBJECT’SINSTANCEATTRIBUTES(Instance=1)
ANALOG
ACTUATOR
OBJECT
ATTRI
BUTE
ATTRIBUTENAME
SERVICE
CODE
DATATYPE
FLOW
BUS
Comment
0x32 0x01 NumberofAttributes 0x0E USINT 7
0x32 0x02 AttributeList 0x0E ArrayofUSINT
0x32 0x03 DataType 0x0E,0x10 USINT 0xC3=INT,0xCA=REAL
0x32 0x04 DataUnits 0x0E,0x10 UINT0x1001=counts,0x1007=%
0x32 0x05 Override 0x0E,0x10 USINT
0x32 0x06 ActuatorValue(valve) 0x0E,0x10 INTorREAL
1
114,1 Seeattribute3and4Fordataunit
countsthevalueattributewillbeinthe
rangeof0..32767.
Where: 0=valveclosed
32767=valvefullyopen
0x32 0x07 Status 0x0E BYTEalways0
0x32 0x15 SafeState 0x0E,0x10 USINT
0x32 0x16 SafeValue 0x0E,0x10 INT 114,6
1
DependsonthevalueassignedtotheDataTypeattribute.Ifthevalueofthisattributeis0xC3,the
selecteddatatypeisInteger,andifthevalueofthisattributeis0xCA,theselecteddatatypewillbethe
IEEE754singleprecisionfloatingpoint.
ANALOGACTUATOROBJECTSERVICES
ANALOGACTUATOROBJECT SERVICECODE SERVICENAME SERVICEDESCRIPTION
0x32 0x0E Get_Attribute_Single Returnsthecontentsofthespecifiedattribute
0x32 0x10 Set_Attribute_Single Modifiesanattributevalue
Noindicationisavailableifcounterlimitisreached.
Aworkaroundcouldbe:
readcounterlimitandcountervalueallthetimebythemasterapplication.
Ifcountervalue>=counterlimit,thebatchhasbeenreached.