4.4.2 Override
TheOverrideattributeisimplementedasfollows:
Override ValuetoValve
0(Normal) setbyattribute6(ActuatorValue)
1(Close) 0
2(Open) max.value
3(Hold) lastusedvalue
4(Safestate) Seesection4.4.3.
4.4.3 Safestate
Inthefollowi ngtabletheimplementedvaluesoftheValveoutputareshown.MappingSafestate:
Safestate ValuetoValve
0(zero/off) 0
1(fullscale/on) max.value
2(holdlastvalue) lastusedvalue
3(usesafevalue) Safevalue(obj0x32,attr0x16)
4.5 GASCALIBRATIONOBJECTINSTANCE
4.5.1 Gasstandardnumber
The retrieval of the gas standard number is implemented as follows: the first 2 characters of the fluid name are
interpretedasadecimalnumber.Thisnumberisreturnedasthegasstandardnumber.
4.5.2 GasSymbol
ThefluidnameisreturnedastheGasSymbol.
4.6 NETWORKERRORDETECTION
WhenanetworkerrorisdetectedwhilethedeviceisinExecutingstate,itissetbacktotheIdleorCriticalFaultstate.
InthosestatesthevalveissettotheSafestate(seeattributeSafeStateinAnalogActuatorobject).Whenthedevice
detectsthatnobuspower
isavailable,thenetworkisclosedandnocommunicationisstarteduntilthebuspoweris
detectedagain.
SucherrorsareshownbytheLED’sontheinstrument.Seethechapter“Troubleshooting”